BRONKHORST HIGH-TECH B.V.
9.17.057 page 8
Whenconnectingthesystemtootherdevices(e.g.toPLC),besurethattheintegrityoftheshieldingis
notaffected.Donotuseunshieldedwireterminals.
Operationviafieldbusisdonebymeansofaflatconductor
cableconnectedwiththemainPCboard.
AlthoughallfunctionalityispossiblebymeansofRS232
andtheswitchontopoftheinstrument,itisimportant
thatcareshouldbetakenwhenremovingtheupperpart
ofthehousing.
Incasethatashutoffvalveismounteddirectlybehindacontrol
module,pleaseapplyarightangledconnectorforAnalog/RS232
communication.
2.7 Analog / Digital operation
2.7aAnalog/Localoperation
ConnecttheMassFloworPressureMeter/Controller
tothepowersupply/readoutunitusingacablewith
9pinsubDconnector.
EXAMPLE
Power:
+15...+24 Vdc
Analog input / output:
0…5Vdc / 0…10Vdc
0…20mA / 4…20mA
2.7bDigital/Fieldbusoperation
ForthisprocedureseedescriptionforRS232ortheapplicable
fieldbusoperation.
RS232connectioncable7.03.366enablestouse(free)
BronkhorsttoolingprogramsforWindows
7.03.366
2.8 Multi-functional switch operation
Usingthe2coloredLEDsandtheswitchontheinstruments,
severalactionscanbemonitoredandstarted.ThegreenLED
isusedforstatusindication.TheredLEDisusedforerrors/
warnings/messages.Theswitchcanbeusedtostartseveral
actions,suchasautozero,restorefactorysettingsandbus‐
initializationactions,ifapplicable.
Seespecificzeroprocedurebelow(10)formoredetails.
Error/warning LED
Status LED
Multifunctional
switch