5 Troubleshooting and service
5.1 General
Foracorrectanalysisoftheproperoperationofaflow/pressuremeterorcontrolleritisrecommendedtoremovethe
unitfromtheprocesslineandcheckitwithoutapplyingfluidsupplypressure.Incasetheunitisdirty,thiscanbe
ascertainedimmediatelybylooseningthecompressiontypecouplingsand, ifapplicabletheflangeontheinletside.
Energizingordeenergizingoftheinstrumentindicateswhetherthereisanelectronicfailure.Afterthat,fluidpressure
istobeappliedinordertocheckbehaviour.Ifthereshouldbesuspicionofleakageincaseofagasunit,donotcheck
forbubbleswithaleakdetectionliquidunderthecoverasthismayleadtoashortcircuitinthesensororp.c.board.
5.2 LED indication
TheredLEDontheinstrumentgiveserrororwarninginformation.
RedLED Time Indication
Off Continuously Noerror
Flash Variable BusactivityontheModbusinterface
On Continuously Criticalerrormessage
Aseriouserroroccurredintheinstrument
Instrumentneedsservicebeforefurtherusing
Formoreinformationchecktheinstructionmanualsfordigitalcommunication/interfacesat
http://www.bronkhorst.com/en/downloads/instruction_manuals/